 |
| Download this
document in PDF format.
PDF reading software can be found here. |
Oxygen Therapy During RTube™ Sampling
Purpose: To provide the same or higher concentration
of oxygen to the patient during breath condensate sampling.
Equipment List
- RTube
- 15 mm I.D./22 mm O.D. multi adaptor
- 6 inches of corrugated tubing
- 22 mm I.D./22 mm O.D. oxygen attachment adapter
- 6 ft. of star lumen oxygen tubing
- Oxygen flow meter 0-15 l/min
- Oxygen source
Procedure
- 1. In order to provide the same FIO2 or higher you will
need to determine the patient's approximate minute ventilation
and current FIO2. Most patients' spontaneous tidal volumes
are 5-7 cc/kg of ideal body weight
- Add the 6 inch reservoir to the RTube by using the 15 mm
I.D./ 22 mm O.D. multi adapter and the oxygen attachment
adapter.
- Place patient on pulse oximeter.
- Use the below FIO2 estimation equation to determine the
approximate oxygen flow required to meet the patient's FIO2
demand.
O2 represents the flow in liter per minute of 100% oxygen
provided to the RTube. The V insp. is the patient's average
inspiratory flow rate, simply multiply the patient's minute
ventilation by the sum of the I:E ratio parts (I:E ratio
being observed or estimated). As an example:
VT
= 200 ml
Respiratory
rate = 16
Min.
ventilation = 3.2 liters/min
Observed
I:E ratio = 1:2
Sum
of I:E parts = 3
Est.
average insp. flow = 9.6 liters/min
Now assume that we are going to add 2 L/min to the RTube
with the above ventilator parameter.
Estimated FIO2 is .37
- Connect the modified RTube to the oxygen flowmeter and
titrate to the appropriate liter flow for estimated FIO2
delivery.
- Remove patient oxygen source and have them breath through
the modified RTube normally.
- Monitor SPO2 and adjust oxygen flow to keep within acceptable
range throughout sampling.
References
- Egan's Fundamentals of Respiratory Care. Scanlan, C;
Spearman, C; Sheldon, R; Fifth Edition 1990
Date Written: 4/7/2004 |